CVE-2014-5431

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Published: 2019-03-26 Last update: 2026-06-16 Assigner: [email protected] Source: [email protected]

Conclusion & alert: CVE-2014-5431 is rated Low Risk (37.8/100): CVSS Medium severity, with low exploitation likelihood (EPSS 0.38%). Mandatory action: Monitor for updates and reassess as exploit intelligence or EPSS changes.

Risk is dynamic; we continuously reassess and refresh what is shown on this page as upstream context changes.

Exploit prediction scoring system (EPSS) score for CVE-2014-5431

EPSS lead: Daily EPSS estimates relative likelihood of exploitation; percentile ranks this CVE among scored vulnerabilities (higher = more severe relative rank).

# Date Old EPSS score New EPSS score Delta (New - Old)
1 2026-06-15 0.05% 0.38% +0.33%
2 2025-03-23 0.05% 0.05% +0.00%
3 2025-03-17 0.05%

Full EPSS history (5 records total)

Common vulnerability scoring system (CVSS) metrics for CVE-2014-5431

CVSS metrics for this CVE.

Base score Version Severity Vector Exploitability Impact Score source
6.8 3.0 MEDIUM
CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H Click to expand
Attack vector (AV:P)
Hands-on access—USB, keyboard, opening the case—not something you do purely over the wire.
Attack complexity (AC:L)
Once they can reach the bug, pulling it off is straightforward—no weird race conditions or rare setup.
Privileges required (PR:N)
No account or special rights needed—anonymous or random user is enough.
User interaction (UI:N)
Nobody has to click “OK” or open a trap file; it can work without a victim helping.
Scope (S:U)
Damage stays in the same “trust bubble” as the broken component—no big spill into unrelated systems.
Confidentiality (C:H)
Serious risk that confidential data gets exposed in a big way.
Integrity (I:H)
They could widely tamper with or forge data—trust in the data is badly hurt.
Availability (A:H)
Could take the service down hard or make it unusable for people who depend on it.
0.9 5.9 [email protected]
4.6 2.0 MEDIUM
AV:L/AC:L/Au:N/C:P/I:P/A:P Click to expand
Access vector (AV:L)
Requires local access to the target system.
Access complexity (AC:L)
Exploitation conditions are straightforward and predictable.
Authentication (AU:N)
No authentication is required.
Confidentiality impact (C:P)
Partial confidentiality impact.
Integrity impact (I:P)
Partial integrity impact.
Availability impact (A:P)
Partial availability impact.
3.9 6.4 [email protected]

Weakness enumeration for CVE-2014-5431

Affected software / configurations for CVE-2014-5431

Vendor Product Version Raw CPE
baxter sigma_spectrum_infusion_system_firmware 6.05 cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*

References for CVE-2014-5431

URL Tags
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 Mitigation Third Party Advisory US Government Resource
cvelogic Threat Intelligence