Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
Conclusion & alert: CVE-2014-5432 is rated Moderate Risk (62.9/100): CVSS Critical severity, with medium exploitation likelihood (EPSS 0.59%). Mandatory action: Review affected assets and schedule remediation.
Risk is dynamic; we continuously reassess and refresh what is shown on this page as upstream context changes.
EPSS lead: Daily EPSS estimates relative likelihood of exploitation; percentile ranks this CVE among scored vulnerabilities (higher = more severe relative rank).
| # | Date | Old EPSS score | New EPSS score | Delta (New - Old) |
|---|---|---|---|---|
| 1 | 2025-08-10 | 0.76% | 0.59% | -0.17% |
| 2 | 2025-03-30 | 1.81% | 0.76% | -1.05% |
| 3 | 2025-03-29 | — | 1.81% | — |
Full EPSS history (8 records total)
CVSS metrics for this CVE.
| Base score | Version | Severity | Vector | Exploitability | Impact | Score source |
|---|---|---|---|---|---|---|
| 9.8 | 3.0 | CRITICAL |
|
3.9 | 5.9 | [email protected] |
| 7.5 | 2.0 | HIGH |
|
10.0 | 6.4 | [email protected] |
| Vendor | Product | Version | Raw CPE |
|---|---|---|---|
| baxter | sigma_spectrum_infusion_system_firmware | 6.05 | cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:* |
| URL | Tags |
|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | Mitigation Third Party Advisory US Government Resource |