Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Conclusion & alert: CVE-2015-3954 is rated Moderate Risk (60.5/100): CVSS Critical severity, with medium exploitation likelihood (EPSS 0.54%). Mandatory action: Review affected assets and schedule remediation.
Risk is dynamic; we continuously reassess and refresh what is shown on this page as upstream context changes.
EPSS lead: Daily EPSS estimates relative likelihood of exploitation; percentile ranks this CVE among scored vulnerabilities (higher = more severe relative rank).
| # | Date | Old EPSS score | New EPSS score | Delta (New - Old) |
|---|---|---|---|---|
| 1 | 2025-03-30 | 1.95% | 0.54% | -1.42% |
| 2 | 2025-03-29 | 0.54% | 1.95% | +1.42% |
| 3 | 2025-03-17 | — | 0.54% | — |
Full EPSS history (5 records total)
CVSS metrics for this CVE.
| Base score | Version | Severity | Vector | Exploitability | Impact | Score source |
|---|---|---|---|---|---|---|
| 9.8 | 3.0 | CRITICAL |
|
3.9 | 5.9 | [email protected] |
| 10.0 | 2.0 | HIGH |
|
10.0 | 10.0 | [email protected] |
| Vendor | Product | Version | Raw CPE |
|---|---|---|---|
| pifzer | plum_a\+_infusion_system_firmware | <= 13.4 | cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:* |
| pifzer | plum_a\+3_infusion_system_firmware | <= 13.6 | cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:* |
| pifzer | symbiq_infusion_system_firmware | <= 3.13 | cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:* |
| URL | Tags |
|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 | Mitigation Third Party Advisory US Government Resource |