Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Conclusion & alert: CVE-2015-3956 is rated Moderate Risk (54.4/100): CVSS Critical severity, with low exploitation likelihood (EPSS 0.24%). Mandatory action: Review affected assets and schedule remediation.
Risk is dynamic; we continuously reassess and refresh what is shown on this page as upstream context changes.
EPSS lead: Daily EPSS estimates relative likelihood of exploitation; percentile ranks this CVE among scored vulnerabilities (higher = more severe relative rank).
| # | Date | Old EPSS score | New EPSS score | Delta (New - Old) |
|---|---|---|---|---|
| 1 | 2025-03-30 | 0.32% | 0.24% | -0.08% |
| 2 | 2025-03-29 | 0.21% | 0.32% | +0.10% |
| 3 | 2023-03-07 | — | 0.21% | — |
Full EPSS history (4 records total)
CVSS metrics for this CVE.
| Base score | Version | Severity | Vector | Exploitability | Impact | Score source |
|---|---|---|---|---|---|---|
| 9.8 | 3.0 | CRITICAL |
|
3.9 | 5.9 | [email protected] |
| 10.0 | 2.0 | HIGH |
|
10.0 | 10.0 | [email protected] |
| Vendor | Product | Version | Raw CPE |
|---|---|---|---|
| pifzer | plum_a\+_infusion_system_firmware | <= 13.4 | cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:* |
| pifzer | plum_a\+3_infusion_system_firmware | <= 13.6 | cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:* |
| pifzer | symbiq_infusion_system_firmware | <= 3.13 | cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:* |
| URL | Tags |
|---|---|
| https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 | Mitigation Third Party Advisory US Government Resource |