GHSA-9p7r-jmp9-7hw2 · Severity: medium — Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling...
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Conclusion & alert: CVE-2019-25723 is rated Low Risk (29.9/100): CVSS Medium severity, with low exploitation likelihood (EPSS 0.04%). Mandatory action: Monitor for updates and reassess as exploit intelligence or EPSS changes.
Risk is dynamic; we continuously reassess and refresh what is shown on this page as upstream context changes.
EPSS lead: Daily EPSS estimates relative likelihood of exploitation; percentile ranks this CVE among scored vulnerabilities (higher = more severe relative rank).
| # | Date | Old EPSS score | New EPSS score | Delta (New - Old) |
|---|---|---|---|---|
| 1 | 2026-06-03 | — | 0.04% | — |
Full EPSS history (1 record total)
CVSS metrics for this CVE.
| Base score | Version | Severity | Vector | Exploitability | Impact | Score source |
|---|---|---|---|---|---|---|
| 6.3 | 4.0 | MEDIUM |
|
— | — | [email protected] |
| 4.0 | 3.1 | MEDIUM |
|
2.2 | 1.4 | [email protected] |
GHSA-9p7r-jmp9-7hw2 · Severity: medium — Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling...
| Vendor | Product | Version | Raw CPE |
|---|---|---|---|
| No affected products in dataset. | |||