Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
Conclusion & alert: CVE-2020-12040 is rated Moderate Risk (59.6/100): CVSS Critical severity, with medium exploitation likelihood (EPSS 0.94%). Mandatory action: Review affected assets and schedule remediation.
Risk is dynamic; we continuously reassess and refresh what is shown on this page as upstream context changes.
EPSS lead: Daily EPSS estimates relative likelihood of exploitation; percentile ranks this CVE among scored vulnerabilities (higher = more severe relative rank).
| # | Date | Old EPSS score | New EPSS score | Delta (New - Old) |
|---|---|---|---|---|
| 1 | 2026-06-15 | 0.18% | 0.94% | +0.76% |
| 2 | 2025-03-17 | 0.14% | 0.18% | +0.04% |
| 3 | 2025-02-12 | — | 0.14% | — |
Full EPSS history (8 records total)
CVSS metrics for this CVE.
| Base score | Version | Severity | Vector | Exploitability | Impact | Score source |
|---|---|---|---|---|---|---|
| 9.8 | 3.1 | CRITICAL |
|
3.9 | 5.9 | [email protected] |
| 5.0 | 2.0 | MEDIUM |
|
10.0 | 2.9 | [email protected] |
| Vendor | Product | Version | Raw CPE |
|---|---|---|---|
| baxter | sigma_spectrum_infusion_system_firmware | >= 6.0, <= 6.05 | cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:*:*:*:*:*:*:*:* |
| baxter | sigma_spectrum_infusion_system_firmware | 8.0 | cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:* |
| URL | Tags |
|---|---|
| https://www.us-cert.gov/ics/advisories/icsma-20-170-04 | Third Party Advisory US Government Resource |