GHSA-v6xg-wr2p-xrj3 · Severity: critical — Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote...
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
Conclusion & alert: CVE-2026-7251 is rated Moderate Risk (46.8/100): CVSS Critical severity, with low exploitation likelihood (EPSS 0.10%). Mandatory action: Review affected assets and schedule remediation.
Risk is dynamic; we continuously reassess and refresh what is shown on this page as upstream context changes.
EPSS lead: Daily EPSS estimates relative likelihood of exploitation; percentile ranks this CVE among scored vulnerabilities (higher = more severe relative rank).
| # | Date | Old EPSS score | New EPSS score | Delta (New - Old) |
|---|---|---|---|---|
| 1 | 2026-05-27 | — | 0.10% | — |
Full EPSS history (1 record total)
CVSS metrics for this CVE.
| Base score | Version | Severity | Vector | Exploitability | Impact | Score source |
|---|---|---|---|---|---|---|
| 9.3 | 4.0 | CRITICAL |
|
— | — | [email protected] |
| 9.8 | 3.1 | CRITICAL |
|
3.9 | 5.9 | [email protected] |
GHSA-v6xg-wr2p-xrj3 · Severity: critical — Eppendorf BioFlo 320 is vulnerable to due to VNC server using a hard-coded password. If a remote...
| Vendor | Product | Version | Raw CPE |
|---|---|---|---|
| No affected products in dataset. | |||