Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
結論預警: CVE-2017-9657 綜合評估為中等風險(44.4/100):CVSS 技術影響為中級,利用機率(EPSS 0.79%) 強制指令: 梳理受影響資產並納入修補計畫。
風險隨態勢動態變化;本站持續評估並同步更新本頁展示內容。
EPSS 日更估計相對被利用可能性;百分位表示該 CVE 在已評分漏洞中的相對排名(越高表示相對更嚴重)。
| # | 日期 | 舊 EPSS 分數 | 新 EPSS 分數 | 變化(新 − 舊) |
|---|---|---|---|---|
| 1 | 2026-06-15 | 0.43% | 0.79% | +0.37% |
| 2 | 2025-05-18 | 0.49% | 0.43% | -0.06% |
| 3 | 2025-03-31 | — | 0.49% | — |
完整 EPSS 歷史 (共 7 筆)
該 CVE 的 CVSS 指標。
| 底座分 | 版本 | 嚴重度 | 向量 | 可利用性 | 影響 | 分數來源 |
|---|---|---|---|---|---|---|
| 6.5 | 3.0 | MEDIUM |
|
2.8 | 3.6 | [email protected] |
| 3.3 | 2.0 | LOW |
|
6.5 | 2.9 | [email protected] |
| 廠商 | 產品 | 版本 | 原始 CPE |
|---|---|---|---|
| philips | intellivue_mx40_firmware | < b.06.18 | cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:* |
| URL | 標籤 |
|---|---|
| http://www.securityfocus.com/bid/100813 | Third Party Advisory VDB Entry |
| https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 | Third Party Advisory US Government Resource |
| https://www.usa.philips.com/healthcare/about/customer-support/product-security | Vendor Advisory |